Sign in
Active, not recruitingNo phase

Study of the RelayPro Thoracic Stent-Graft in Subjects With an Acute, Complicated Type B Aortic Dissection

Key facts

Who runs it
Bolton Medical
Where
Ann Arbor, Atlanta, Aurora, Baltimore, Buffalo, Chicago, Cleveland, Hartford, Indianapolis, Iowa City, Los Angeles, Milwaukee, New Haven, Norfolk, Philadelphia, Portland, Roanoke, Seattle, Washington D.C., and Wynnewood, United StatesSee the 21 sites
Trial code
Status
Active, not recruiting
People taking part
56 people enrolled
Started
14 Dec 2017
Expected to end
Nov 2026
Type of study
Interventional
Treatments tested
  • Relay Pro Stent-Graft
Record updated
6 November 2025

Who can take part

Age
18 years and older
Sex
Women and men
Healthy volunteers
Cannot take part

The registry criteria, word for word

Show all 32 criteriaShow less

Who can take part

  • Subject must have an acute (symptom onset to diagnosis within 2 weeks)or subacute, complicated type B aortic dissection (entire dissection is distal to the left subclavian artery (LSA)), confirmed by Computed Tomography Angiography (CTA) or Magnetic Resonance Angiogram (MRA), with time from symptom onset to diagnosis ≤ 6 weeks, with at least one of the following:; Malperfusion of the viscera, kidneys, spinal cord, or lower extremities, measured by clinical or radiographic evidence; Rupture; Intractable pain
  • Proximal and distal aortic neck with diameter between 19 mm and 42 mm
  • Subject's anatomy must meet all of the following anatomical criteria:; Proximal attachment zone distal to the left common carotid and a distal attachment zone proximal to the origin of the celiac artery. (Dissection is permitted in the distal attachment zone but is not permitted in the proximal attachment zone.); The length of the attachment zones will depend on the intended stent-graft diameter and type of graft selected; The proximal attachment zone should be 15 mm for 22 - 28 mm RelayPro grafts with bare stent (20 mm for RelayPro grafts with non-bare stent), 20 mm for 30 - 46 mm RelayPro grafts with bare stent (25 mm for RelayPro grafts with non-bare stent), and proximal to non-dissected segment (healthy zone); The distal attachment zone should be 20 mm for all RelayPro grafts; Coverage of the left subclavian artery is permitted with mandatory revascularization if patent left internal mammary artery (LIMA) bypass or left upper extremity (LUE) arteriovenous graft or anomalous vertebral artery off the aorta. Revascularization must be performed prior to device placement, and may occur during implant procedure, provided it is before coverage of the LSA by the endograft
  • Proximal attachment zone containing a straight segment (non-tapered, non-reverse-tapered, defined by <10% diameter change) with lengths equal to or greater than the required attachment length for the intended device
  • Vascular dimensions (e.g., aortic diameters, length from left subclavian to celiac artery) must be in the range that can be safely treated with the RelayPro Thoracic Stent-Grafts
  • Adequate iliac or femoral artery access for introduction of the RelayPro Delivery System. Alternative methods to gain proper access may be utilized (e.g., iliac conduit)
  • Subject willing to comply with the follow-up evaluation schedule
  • Subject (or Legally Authorized Representative, LAR) agrees to sign an Informed Consent Form prior to treatment

Who cannot take part

  • Prior endovascular or surgical repair in the descending thoracic aorta. The device may not be placed within any prior endovascular or surgical graft
  • Concomitant aneurysm/disease of the ascending aorta, aortic arch, or abdominal aorta, requiring repair. Dissection extension into the abdominal aorta is acceptable
  • Prior abdominal aortic aneurysm repair (endovascular or surgical) that was performed less than 6 months prior to the planned stent implant procedure
  • Major surgical or medical procedure within 30 days prior to the planned procedure, or is scheduled for a major surgical or medical procedure within 30 days post implantation. This excludes any planned procedures for the prospective stent-graft placement
  • Acute renal failure or chronic renal insufficiency, and not receiving dialysis
  • Current or planned participation in an investigational drug or device study that has not completed primary endpoint evaluation
  • Untreatable allergy or sensitivity to contrast media or device components, including metal stents
  • Known or suspected connective tissue disorder
  • Blood coagulation disorder or bleeding diathesis for which the treatment cannot be suspended for one week pre- and/or post-repair
  • Coronary artery disease with unstable angina
  • Severe congestive heart failure (New York Heart Association functional class IV)
  • Stroke and/or Myocardial Infarction (MI) within 3 months of the planned treatment date
  • Pulmonary disease requiring the routine (daily or nightly) need for oxygen therapy outside the hospital setting
  • Active systemic infection and/or mycotic aneurysms
  • Morbid obesity or other condition that may compromise or prevent the necessary imaging requirements
  • Diagnosis of traumatic injury or transection of the descending thoracic aorta
  • Significant stenosis, calcification, thrombus, or tortuosity of intended fixation sites that would compromise fixation or seal of the device
  • Planned coverage of left carotid or celiac arteries; or anatomic variants that would compromise circulation to the carotid, vertebral, or innominate arteries after device placement, which is not amenable to subclavian revascularization
  • Hemodynamically unstable
  • Bowel necrosis
  • ASA risk classification = V (Moribund patient not expected to live 24 hours with or without operation)
  • Less than two-year life expectancy
  • Currently pregnant or planning to become pregnant during the course of the study
  • Medical, social, or psychological issues that Investigator believes may interfere with treatment or follow-up

Where in United States

  • University of Michigan Health SystemAnn Arbor, United States
  • Emory UniversityAtlanta, United States
  • University of Colorado DenverAurora, United States
  • University of Maryland Medical Center (UMB)Baltimore, United States
  • Buffalo General Medical CenterBuffalo, United States
  • Northwestern Memorial HospitalChicago, United States
  • Cleveland Clinic FoundationCleveland, United States
  • University HospitalsCleveland, United States
  • Hartford HospitalHartford, United States
  • St. Vincent Heart CenterIndianapolis, United States
  • University of Iowa Hospital and ClinicIowa City, United States
  • USC Department of SurgeryLos Angeles, United States
  • Medical College of WI, Vascular SurgeryMilwaukee, United States
  • Yale School of MedicineNew Haven, United States
  • Sentara Heart HospitalNorfolk, United States
  • University of Pennsylvania Medical Center / Penn PresbyterianPhiladelphia, United States
  • Providence Heart InstitutePortland, United States
  • Carilion Roanoke Memorial HospitalRoanoke, United States
  • University of WashingtonSeattle, United States
  • MedStar Washington Hospital CenterWashington D.C., United States
  • Lankenau Medical CenterWynnewood, United States

About the trial

What the trial wants to find out

This clinical trial is a prospective, multicenter, non-blinded, non-randomized study designed to assess the RelayPro thoracic endografts in the treatment of acute, complicated type B aortic dissection. The primary endpoint will measure all-cause mortality at 30 days post-procedure.

The group in the trial

  • Experimental: Relay ProTreatment being tested

    The Relay Pro arm includes subjects who receive the device. The Relay Pro Stent Graft System is administered to treat complicated Type B aortic dissections.

    Treatment: Relay Pro Stent-Graft

Data and sources

Registry record
Listed on ClinicalTrials.gov.How to verify a trial in the registry
Official title
A Prospective, Multicenter, Non-Blinded, Non-Randomized Study of the RelayPro Thoracic Stent-Graft in Subjects With an Acute, Complicated Type B Aortic Dissection
Record updated