Pulse oximeter accuracy and skin color in children with congenital heart disease
Measures how skin pigment affects pulse oximeter readings in people with congenital heart disease.
Key facts
- Who runs it
- Icahn School of Medicine at Mount Sinai
- Where
- New York, United StatesSee the site
- Trial code
- Registry title
- Skin Pigment/Pulse Oximeter in Congenital Heart Disease (CHD)
- Status
- Recruiting
- People taking part
- 24 people, planned
- Started
- 29 Oct 2024
- Expected to end
- 31 Aug 2027
- Type of study
- Observational
- Record updated
- 25 September 2026
Who can take part
AI summary
From the trial's eligibility criteria. Machine-read from the registry text on this page, so it can be wrong. The registry text below and the trial team decide who can take part.
The trial is looking for children with congenital heart disease (a heart problem present from birth) who are having heart surgery. It checks how accurate pulse oximeters (finger oxygen sensors) are across different skin colours.
- Age
- Under 18 years old
- Sex
- All sexes
- Disease stage
- Congenital heart disease, cyanotic (causing low blood oxygen, bluish skin) or acyanotic. Children must be having heart surgery under general anaesthesia, with a planned line into an artery.
Main reasons for not taking part
- Age greater than 18 years old
- Emergency surgery
- Significant anaemia before surgery (hemoglobin <8.0 g/dL)
- Unstable heart and blood flow before surgery, such as more than 1 drip of heart-supporting medicine or a machine supporting circulation
- Coloured nail polish where the sensor goes, planned dyes given into a vein during surgery, or refusal by the patient, parent or guardian
The registry criteria, word for word
Show all 6 criteriaShow less
Who can take part
- Pediatric patients (age less than 18 years old) with a diagnosis of CHD (cyanotic or acyanotic) who are presenting for cardiac surgery under general anesthesia with planned arterial access
Who cannot take part
- Emergency surgery
- Significant preoperative anemia (hemoglobin <8.0 g/dL)
- Age greater than 18 years old
- Preoperative hemodynamic instability (i.e., >1 vasoactive infusions or mechanical circulatory support)
- The presence of any colored nail polish on the planned site of pulse oximeter placement, planned use of any intravenous dyes intraoperatively, and patient, parent, or guardian refusal
Where in United States
- Mount Sinai HospitalNew York, United StatesContact: Garrett W. Burnett, M.D.Taking people
About the trial
What the trial wants to find out
Recent retrospective studies have demonstrated differences between pulse oximeter values (SpO2) and measured arterial oxygen saturation (SaO2) in patients identifying as Black or Hispanic. These retrospective studies have limitations because self-reported race is likely not an accurate metric for level of skin pigmentation and the retrospective nature of these studies may impact the accuracy of simultaneous measures of arterial oxygen saturation and pulse oximeter values. The few prospective studies that have evaluated this issue have utilized color-matching techniques to quantify skin pigmentation, and fewer studies have directly measured skin pigmentation in relation it to pulse oximeter accuracy. The aim of this study is to prospectively measure pulse oximeter accuracy in relation to measured levels of skin pigmentation in the congenital heart disease population.
The group in the trial
- Pediatric Patients with Congenital Heart Disease
No intervention will take place in this observational study. All patients will have their skin pigment measured using a color spectrophotometer and color matching techniques (Fitzpatrick Scale, Monk Skin Tone Scale). Two pulse oximeters will be utilized and pulse oximeter readings will be compared to simultaneous measured arterial saturations. After cardiopulmonary bypass is initiated for the surgery, the subject's involvement in the study will be completed.
Questions to ask the study team
- How often do I need to come in, and for how long each time?
- Are travel costs paid?
- Can I keep my own doctor?
- What happens if I want to stop?
Contact the study team: see the registry record → Contacts and Locations
Print a page for my doctorSaves a reminder to your own calendar to look at this page again. Frontlight sends no messages and does not know you set it.
Data and sources
- Registry record
- Listed on ClinicalTrials.gov.How to verify a trial in the registry
- Official title
- Pulse Oximetry Accuracy and Skin Pigmentation in Congenital Heart Disease: A Prospective Observational Study
- Plain-language title
- Pulse oximeter accuracy and skin color in children with congenital heart diseaseGenerated from the registry record. The registry title is the official one.
- AI summary
- From the trial's eligibility criteria. Read by machine on 14 November 2026.
- Record updated