Sign in
CompletedNo phase

Use of AlignRT System During Stereotactic Radiosurgery

Key facts

Who runs it
Institute of Oncology Ljubljana
Where
Slovenia1 site in 1 countryChoose a country to see its sites
Trial code
Status
Completed
People taking part
20 people enrolled
Started
1 Apr 2025
Ended
31 Aug 2025
Type of study
Interventional
Treatments tested
  • Manual Patient Positioning
  • AlignRT-Guided Patient Positioning
Record updated
21 July 2026

Who can take part

Age
18 years and older
Sex
Women and men
Healthy volunteers
Cannot take part
Show the full registry criteria (7)Hide the full registry criteria

Who can take part

  • Adults aged 18 years or older
  • Patients with brain metastases eligible for single-fraction stereotactic radiosurgery
  • Treatment performed using the TrueBeam STx linear accelerator
  • Able and willing to provide written informed consent

Who cannot take part

  • Patients receiving fractionated stereotactic radiotherapy
  • Patients younger than 18 years
  • Treatment performed using a linear accelerator other than TrueBeam STx

Where is this trial?

It has sites in one country. Choose a country to see its sites.

About the trial

What the trial wants to find out

This prospective study evaluated whether the AlignRT surface-guided radiotherapy system improves patient positioning accuracy during stereotactic radiosurgery (SRS) for brain metastases compared with standard manual positioning. Twenty adults scheduled for single-fraction SRS were positioned twice before treatment. Participants were first positioned using the standard manual method with delta couch shift and were then positioned automatically using the AlignRT system. After each positioning method, a partial cone-beam computed tomography (CBCT) scan was performed to measure residual setup deviations in three translational and three rotational directions. The primary objective was to compare the average setup deviations between the two positioning methods and determine whether AlignRT provided more accurate patient positioning. The additional positioning and imaging procedure prolonged the treatment visit by up to 15 minutes and involved a low additional imaging dose.

The group in the trial

  • Manual and AlignRT-Guided PositioningTreatment being tested

    Participants underwent standard manual positioning followed by automatic positioning using the AlignRT system before stereotactic radiosurgery. A partial cone-beam computed tomography scan was performed after each positioning method to measure residual translational and rotational setup deviations.

    Treatment: Manual Patient Positioning and AlignRT-Guided Patient Positioning

Data and sources

Registry record
Listed on ClinicalTrials.gov.How to verify a trial in the registry
Record updated
21 July 2026